18 User Manual
Traveling with the System
When traveling, the carrying case is for carry-on luggage only. The carrying case will not protect the system if it is put
through checked baggage.
For your convenience at security stations, there is a note on the bottom of the device stating that it is medical
equipment and is suitable for airline use. It may be helpful to bring this manual along with you to help security
personnel understand the BiPAP autoSV Advanced System One device.
If you are traveling to a country with a line voltage different than the one you are currently using, a different power
cord or an international plug adaptor may be required to make your power cord compatible with the power outlets
of the country to which you are traveling. Contact your home care provider for additional information.
• Airline Travel
The device is suitable for use on airlines when the device is operating from an AC or DC power source.
Home Cleaning
• Cleaning the Device
WARNING: To avoid electrical shock, always unplug the power cord from the wall outlet before cleaning the
device. DO NOT immerse the device in any uids.
1. Unplug the device, and wipe the outside of the device with a cloth slightly dampened with water and a mild
detergent. Let the device dry completely before plugging in the power cord.
2. Inspect the device and all circuit parts for damage after cleaning. Replace any damaged parts.
• Cleaning or Replacing the Filters
Under normal usage, you should clean the gray foam lter at least once every two weeks and replace it with a
new one every six months. The white ultra-ne lter is disposable and should be replaced after 30 nights of use or
sooner if it appears dirty. DO NOT clean the ultra-ne lter.
CAUTION: Dirty inlet lters may cause high operating temperatures that may affect device performance.
Regularly examine the inlet lters as needed for integrity and cleanliness.
1. If the device is operating, stop the airow. Disconnect the device from the power source.
2. Remove the lter(s) from the enclosure by gently squeezing the lter in the center and pulling it away from the
device.
3. Examine the lter(s) for cleanliness and integrity.
4. Wash the gray foam lter in warm water with a mild detergent. Rinse thoroughly to remove all detergent
residue. Allow the lter to air dry completely before reinstalling it. If the foam lter is torn, replace it. (Only
Respironics-supplied lters should be used as replacement lters.)
5. If the white ultra-ne lter is dirty or torn, replace it.
6. Reinstall the lters, inserting the white ultra-ne lter rst if applicable.
CAUTION: Never install a wet lter into the device. You must ensure sufcient drying time for the cleaned lter.
• Cleaning the Tubing
Clean the tubing before rst use and daily. Disconnect the exible tubing from the device. Gently wash the tubing
in a solution of warm water and a mild detergent. Rinse thoroughly. Air dry.
Service
The device does not require routine servicing.
WARNING: If you notice any unexplained changes in the performance of this device, if it is making unusual or
harsh sounds, if it has been dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is
broken, disconnect the power cord and discontinue use. Contact your home care provider.