31
Troubleshooting
Problem Possible Cause Possible Solution
The device won’t
turn on.
The batteries are
depleted.
Replace all 4 batteries.
The batteries are
installed incorrectly.
Verify battery
orientation, illustrated
inside the battery
compartment or in
Figure 3: Installing
Batteries section of
this operators manual.
A metal contact in the
battery compartment is
missing or damaged.
Contact Respironics
Technical Service.
A dash appears in
the leftmost
position of the
SpO
2
display.
A sensor fault exists
(disconnect, failure,
misalignment, or
incompatibility with the
monitor).
Verify that the sensor
is correctly connected
to the device and the
patient; replace sensor
if the condition
persists.
Dashes are
displayed in both
the SpO
2
and Pulse
Rate displays.
No signal is detected
because the sensor is
not plugged in.
Verify the sensor
connections.
Sensor failure. Replace the sensor.